Use the ARROW keys to set the reading on the secondary LCD to that
of the hot water.
When the reading is stable and close to the
selected calibration point, “READY” tag will
appear and “CFM” tag will blink.
Press CFM to confirm. The instrument returns
to measurement mode.
Note: If the reading is not close to the selected calibration point,
WRONG” tag will blink. Change the temperature probe and
restart calibration.