13
The   tag will blink on the LCD until the reading is
stable.
When the reading is stable and close to the selected buffer,
READY tag will be displayed and CFM tag will blink.
Press CFM to confirm calibration.
The calibrated value will be displayed on the primary LCD
and the second expected buffer value on the secondary
LCD line.
Note: The instrument will automatically skip the buffer used for the first point.
It also skips 6.86 if 7.01 buffer was used and viceversa. Likewise, it will
skip 9.18 if 10.01 buffer was used and viceversa.
After the first calibration point is confirmed, immerse the pH electrode and
the temperature probe approximately 4 cm (1½) into the second buffer
solution and stir gently. The temperature probe should be close to the pH
electrode.
If necessary, press the ARROW keys to select a different buffer value.
The   tag will blink on the LCD until the reading is
stable.
When the reading is stable and close to the selected buffer,
READY tag will be displayed and CFM tag will blink.
Press CFM to confirm calibration.
The calibrated value will be displayed on the primary LCD
and the third expected buffer value on the secondary LCD
line.
Note: The instrument will automatically skip the buffer used for the first and
second calibration point.
After the second calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½) into the third
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The   tag will blink on the LCD until the reading is
stable.
When the reading is stable and close to the selected buffer,
READY tag will be displayed and CFM tag will blink.
Press CFM to confirm calibration.
The instrument stores the calibration value and returns
to normal measurement mode.
BUFFER pH
2
BUF
pHCAL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 255 and is the answer not in the manual?

Hanna Instruments HI 255 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 255
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals