Immerse the temperature probe (or EC probe) in the second vessel as near
as possible to the calibrated thermometer. Allow a few seconds for the
probe to stabilize.
Use the ARROW keys to set the reading on the secondary LCD to that of
the hot water.
When the reading is stable and close to the selected
calibration point, READY tag will appear and CFM
tag will blink.
Press CFM to confirm. The instrument returns to
measurement mode.
Note: If the reading is not close to the selected calibration point, WRONG
tag will blink. Change the temperature probe (or EC probe) and restart
calibration.