28
The temperature compensation mode.
The instrument ID.
Notes: If no temperature compensation is selected during calibration,
the temperature coefficient is not displayed in GLP.
Press GLP at any moment and the instrument will return to
measurement mode.
If calibration has not been performed on the selected range,
the instrument displays “no CAL” message blinking.
CAL
LOGGING FUNCTIONLOGGING FUNCTION
LOGGING FUNCTIONLOGGING FUNCTION
LOGGING FUNCTION
Up to 700 logged samples can be stored into memory.
200 manually logged records and 500 lot logging records can be stored
in the memory. To select logging type enter SETUP menu.
LOGGING THE CURRENT DATA LOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA LOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA
(manual logging)(manual logging)
(manual logging)(manual logging)
(manual logging)
Select the manual logging mode in SETUP menu.
To store the current reading into memory press LOG while the instrument
is in measurement mode.
The instrument will display “MAn” on the primary LCD, the record
number on the secondary LCD and “LOG” tag for a few seconds (see
example below: record No. 15):