EasyManuals Logo

Hanna Instruments HI 8711 User Manual

Hanna Instruments HI 8711
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
31
SPECIFICATIONS HI 8720
HI 931500
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy ±0.02 pH
Typical EMC ±0.1 pH
Deviation ±0.2 mA
Installation Cat. II
Input 10
12
Ohm
Calibration Offset: ±2 pH O trimmer
Slope: 80 to 110% Slope trimmer
Temperature Fixed or automatic with PT100
Compensation from 0 to 100°C
Readout 4-digit LCD plus graphic symbols
Recorder Output 0 to 20 mA or 4 to 20 mA
Set Point Relay One, Isolated, 2 A, max 240 V,
resistive load, 1.000.000 strokes
Power Supply 110/115 V or 220/240 V; 50/60 Hz
Environment -10 to 50°C (14 to 122°F)
Weight 1 kg (2.2 lb.)
SPECIFICATIONS HI 931500
HI 8720E HI 8720T
Range -1000 to +1000 mV
Resolution 1 mV
Accuracy ±5 mV ±0.5%
Typical EMC ±6 mV ±6 mV
Deviation ±0.2 mA ±0.2 mA
Installation Cat. II
Input 10
12
Ohm 4 to 20 mA
Calibration Slope: 90 to 110% Slope trimmer
Set Point Relay One, Isolated, 2 A, max 240 V,
resistive load, 1.000.000 strokes
Alarm Relay One, Isolated, 2 A, max 240 V,
resistive load, 1.000.000 strokes
Readout 4-digit LCD plus graphic symbols
Recorder Output 0 to 20 mA or 4 to 20 mA (isolated)
Power Supply 110/115 V or 220/240 V; 50/60 Hz
Environment -10 to 50°C (14 to 122°F)
Weight 1 kg (2.2 lb.)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 8711 and is the answer not in the manual?

Hanna Instruments HI 8711 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 8711
CategoryController
LanguageEnglish

Related product manuals