29
Press GLP s to enter or exit GLP data consulting.
Several functions are available when in GLP menu.
Press READ u to scroll the following GLP data:
The last calibration date, in YYYY.MM.DD
format. If no calibration was performed, the
factory calibration message, “F.CAL”, will be
displayed on the LCD.
The time of the last calibration in hh:mm
format.
First calibration point: 0.00 NTU if skipped or
the actual read value (e.g. 0.01 NTU).
Second calibration point.
GOOD LABORATORY PRACTICE (GLP)
The instrument asks for confirmation and if
LOG/CFM is pressed, all records are deleted.
To abort the delete function, press READ u
instead of LOG/CFM.
After all records are deleted, dashes are
displayed for one second and the instrument
returns to idle mode.
The GLP feature allows the user to view last calibration data. Also, the user calibration can be
deleted.