SETUP
To enter/exit SETUP, press SETUP/DEL.
To select the parameter to be edit, press s or
t keys until the desired panel is displayed.
SET EPA COMPLIANCE MODE
When EPA compliance reading is ON, “EPA
message is displayed on the secondary LCD and
the reported values are rounded to meet EPA
reporting requirements.
To edit the EPA mode, press CAL when “EPA
compliance reading panel is displayed. The
parameter setting and “CFM” tag will start
blinking.
Press the s or t keys to set ON or OFF the
EPA compliance mode.
Press LOG/CFM to save the setting. The new
selected option of the parameter will be
displayed on the LCD. Alternatively, press CAL
to exit without saving the new settings.
SET BEEPER
The HI 98703 has a built-in beeper that signals
the tag read, the key press and the error conditions.
The beeper can be selected to be ON or OFF.
To set the beeper ON/OFF, press CAL when
set beeper panel is displayed. The beeper status
and “CFM” tag will start blinking.
Setup mode allows viewing and modifying the instrument parameters.
The blinking “CAL” tag during setup mode suggest to press CAL for parameters editing.
www. .com
information@itm.com1.800.561.8187

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 98703 and is the answer not in the manual?

Hanna Instruments HI 98703 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 98703
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals