EasyManuals Logo

Hanna Instruments HI 98713 User Manual

Hanna Instruments HI 98713
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
18
HEATING THE SAMPLE
Use of heat to remove air bubbles, although very effective in some cases, should be handled with care
because it can alter the turbidity of the sample. When heating a sample, the volatile components
from the sample can vaporize, the suspended components can dissolve or the sample characteristics
can change.
Therefore, the heating procedure should be used with extreme care.
The best way is to use a warm water bath and immerse the cuvet with sample into the bath. Heat the
sample only until the visible bubbles are removed.
Note: Always cool the heated sample to the original sample temperature before measurement.
The heating procedure can be used in combination with vacuum or ultrasonic waves application for
a more effective air bubble removal.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 98713 and is the answer not in the manual?

Hanna Instruments HI 98713 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 98713
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals