If the minimum start DO condition is met the instrument will display the instantaneous specific oxygen
consumption rate and the amount of time that has passed from the beginning of the test.
In case that the SOUR value is corrected to 20 °C (68 °F) and the measured temperature isn’t in the
range 10 to 30 °C the temperature value will blink to alert that the temperature correction isn’t valid.
If the DO reading is less than the minimum end DO value set during SOUR configuration, a warning
icon will be displayed and a beep will be heard every two seconds. Press Stop to stop the test and
the beeper.
To end the SOUR test before the maximum interval set during SOUR configuration press Stop.
If Stop is pressed before the minimum time for the SOUR test has elapsed the instrument will display
a warning message.
SOUR MEASUREMENT