55
PAR Requests the setup parameters setting.
The answer string contains:
• backlight value (1 char)
• contrast value (2 chars)
• instrument ID (4 chars)
• calibration alarm timeout (2 chars)
• Setup information (2 chars): 8 bit hexadecimal encoding:
• 0x01: ‑ beep is on
• 0x04: ‑ Celsius / Fahrenheit display (°C if the bit is set)
• 0x10: ‑ manual pressure (1 if activated, 0 otherwise)
• Auto Light Off time (3 chars)
• Auto Power Off time (3 chars)
• Salinity value (3 chars)
• Pressure unit (1 char): 0 ‑ mmHg, 1 ‑ inHg, 2 ‑ atm, 3 ‑ mbar, 4 ‑ psi, 5 ‑ kPa
• BOD configuration values
• sample minimum delta DO, including sign and decimal point (6 chars)
• sample minimum end DO, including sign and decimal point (6 chars)
• seed minimum delta DO, including sign and decimal point (6 chars)
• seed minimum end DO, including sign and decimal point (6 chars)
• OUR configuration values
• minimum time in seconds (4 chars)
• maximum time in seconds (4 chars)
• minimum start DO including sign and decimal point (6 chars)
• minimum end DO including sign and decimal point (6 chars)
• total volume including sign and decimal point (6 chars)
• sample volume including sign and decimal point (6 chars)
• SOUR configuration values
• minimum time in seconds (4 chars)
• maximum time in seconds (4 chars)
• minimum start DO including sign and decimal point (6 chars)
• minimum end DO including sign and decimal point (6 chars)
• total volume including sign and decimal point (6 chars)
• sample volume including sign and decimal point (6 chars)
• solids weight including sign and decimal point (6 chars)
• SOUR temperature correction (1=enabled, 0=disabled) (1 char)
• The short name of the selected language (3 chars)
PC INTERFACE