NSLx Requests the number of logged samples (4 chars)
The command parameter (x ‑ 1 char)
D ‑ the request is for DO
B ‑ the request is for BOD
O ‑ the request is for OUR
S ‑ the request is for SOUR
I ‑ the request is for BOD initial values
LODDxxx Requests the xxx
th
DO logged data
LODBxxx Requests the xxx
th
BOD logged data
LODOxxx Requests the xxx
th
OUR logged data
LODSxxx Requests the xxx
th
SOUR logged data
LODIxxx Requests the xxx
th
initial BOD value logged data
LODDALL Requests all the DO log on demand
LODBALL Requests all the BOD log on demand
LODOALL Requests all the OUR log on demand
LODSALL Requests all the SOUR log on demand
LODIALL Requests all the initial BOD values log
The answer string for each record contains:
The logged mode (2 chars)
20 ‑ DO range
21 ‑ BOD range
22 ‑ OUR range
23 ‑ SOUR range
24 ‑ BOD initial values
DO log data:
Measurement unit (0 = %, 1 = mg/L) (1 char)
DO value including sign and decimal point (8 chars)
Salinity value [g/L] (3 chars)
Pressure value in mmHg, including sign and decimal point (11 chars)
Temperature value including sign and decimal point (8 chars)
BOD log data:
Sample type (1 = sample, 0 = seed) (1 char)
Seed corrected (1 = corrected, 0 = not corrected) (1 char)
Bottle ID (4 chars)
BOD value including sign and decimal point [mg/L] (8 chars)
Bottle volume including sign and decimal point [ml] (6 chars)
PC INTERFACE
www.tools.in.th
www.neonics.co.th

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98193 and is the answer not in the manual?

Hanna Instruments HI98193 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98193
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals