17. Trouble Shooting
17.1 Potential faults
The table below has been created to assist with res
olving faults you may come across; if the
instructions here do not resolve the issue or the i
ssue you are experiencing is not listed please
contact Ultimate Healthcare using the details found
on the back cover of this manual.
Problem Possible Cause Resolution
LED glows when the
handset is locked
LED does not glow when
the handset is unlocked
Fault in the handset
Fault in the controller
Contact Ultimate
Healthcare using the
details on the back cover
of this manual
No adjustment of the bed
can be made when the
power is ON
Handset is locked (after a
power failure the handset
automatically switches to
the locked state)
Release the handset using
the magnet
The function-LED does
not glow when a key is
pressed
Check the handset is
unlocked
Control unit or handset
defective
Contact Ultimate
Healthcare using the
details on the back cover
of this manual
The handset doesn’t
function – no power light
Mains power not
connected
Handset not connected
to the controller
Connect to the mains
power
Connect the handset to
the controller; following the
colour coding
When a button is pressed
on the handset, the
corresponding actuator
does not respond
Connectors not plugged
into the correct sockets
Faulty actuator
Check the connections,
these are colour coded
Contact Ultimate
Healthcare using the
details on the back cover
of this manual
An unexpected operation
occurs when a button is
pressed on the handset
Connectors not plugged
into the correct sockets
Check the connections,
these are colour coded
Contact Ultimate
Healthcare using the
details on the back cover
of this manual
18.
Reuse
The
Apex Health Care
3URGXFWV
beds are designed for reuse. They must be cleaned and
disinfected prior
to reuseas per Section 15.
19.
Disposal
Beds consist of:
Metal parts
Plastic parts
Electronic components
Electric cables
Rechargeable batteries/batteries