- 13 -
3.5 Drainage
Forsterilizerwithinternalwatertank.Afterusingitafewtimespleasetakeoutplasticdrainconnectorand
pipefromaccessorybagwhentodischargethewastewatertank.Pleaseinsertoneendofthedrainconnector
withoutablackrubbersealintothedrainpipeandinserttheotherintodrainageinterface(Figure1.4andFigure
1.5).
For sterilizer with internal water tank, please empty the waste
watertankbeforeinjectingthedistilledwater.
3.6 Sterilizationeffectverification
Smallsteamsterilizerinstalledorusedfor12monthsoraftermaintenanceshallcarryoutB&Dtestwithout
loadandcarryoutphysical,chemicalandbiologicalmonitoringwithfullload.Aftermonitoringisprovedsuccessful,
thesterilizercanbeusedagain.VerificationofsterilizationresultsshallcomplywiththerequirementsofAppendix
EofWS506‐2016
4 Sterilizationprogram
4.1 Loadstobesterilized
Properloadingisakeyfactorforeffectivesterilizationanddrying.You
mustalsopayattentiontothefollowingwhenloading
The load must be sterilized by the manufacturer if steam
sterilizationispermitted.
Packaging materials for packaged sterilization products shall
complywithGB/T19974standard.
Sterilizeditemspackedinpaperandplasticbags,Thepapermust
befacedupandlaidflatinthetray.
Formixedloadforsterilization,fabricisonthetopandthemetal
isonthelowerlayer.
Theremustbeagapbetweenthesterilizationbags.
Metalbagsanddressingbagscannotbeattachedtogether.
Theloadcapacitymustnotexceed90%ofthechambercapacity.
4.2 Selectsterilizationprogram
This product has 7 preset sterilization programs. Please select the appropriate sterilization program
accordingtotheloadtypeandpackagingmaterial.Theparametersettingofthesterilizationprogramandthetypical
loaddescriptionareasfollows.ThemaximumloadofQ50B/Q51Bis6kginstrumentsor2kgfabric.Themaximum
loadofQ70/Q71Bis7kginstrumentsor2.5kgfabric.