- 15 -
The chamber will automatically heat up to 110 °C when the automatic
preheatsettingisturnedon.Pleasebecarefulofburns!
4.3.2 Automaticprinting
Turnontheautomaticprintingfunctionandthesterilizerisproperlyconnectedtotheprinterto
automaticallyprintthesterilizationrecordaftertheprogramends.Seesection6.5fordetails
4.3.3 Extradrying
Additionaldryingsettingsbasedonuserpreferences,extendthedryingtimeofthepresetsterilization
programbasedonthedryingtimeinTable4‐1toachievegooddryingresultsforloads,especiallycottonfabrics.
Seesection6.6fordetails.
4.4 Runningaprogram
4.4.1 Startsterilization
Afterselectingsterilizationprogram,pleaespress tostartupprogram.Sterilizerwillensurethatthedoor
islockedandthewaterqualityofthedistilledwaterismonitoredafterthestartoftheprogram.
Fig4.2startingprogram
4.4.2 Programrunningprocess
Sterilizationprogramincludes3phases,pre‐vacuum,sterilizationanddrying.Thedescriptionofeachstageis
detailedinsection1.7.1ofthismanual.Thecontrollermonitorstheoperationofthesterilizerandthe
temperatureandpressureinthechamberinrealtimeanddisplaysrelevantinformationonthescreen.
07/25/201709:23:58
134program
134/2.1Bar/5Min
T:25.0P:0.00Bar
Doorstatus
Press to
startprogram