- 30 -
Solution:
Pleasecheckifthefootpadadjustmentiscorrect.Undernormalcircumstances,thefrontsideis
slightlyhigherthanthebackside.
Openthedoorandcheckifthefilteratthebottomofthesterilizationchamberisblocked.
Noneoftheabovecanhelpreducetheconsumptionofdistilledwater,pleasecontactprofessional
distributorauthorizedbyHӓTmed.
9.3 Thedryingeffectisnotgood
Faultydescription:Dryingcannotachievesatisfactoryresults
Solution:
Pleasecheckifthefootpadadjustmentiscorrect.Undernormalcircumstances,thefrontsideis
slightlyhigherthanthebackside.
Openthedoorandcheckifthefilteratthebottomofthesterilizationchamberisblocked.
Pleasecheckiftheloadexceedsthemaximumloadingcapacity(especiallytextiles)andiftheload
isplacedcorrectlyinthechamber(donottouchthewallofthepressurechamberdirectly)andifthe
standardtraysandtrayrackareusedforloading.
Activateautomaticwarm‐up.
Opentheleftsidecoverandcheckifthefilterisclogged
Noneoftheabovecanhelpimprovetheeffectofdrying,pleasecontactprofessionaldistributor
authorizedbyHӓTmed.
9.4 Nodistilledwater
Faultydescription:Nodistilledwater
Solution:
Pleasecheckifthereiswaterindistilledwatertank.
Pleasecheckifthedistilledwaterpipeisbentorknottedandinsertedintoplace.
9.5 Unabletoopenthedoor
Faultydescription:Thedoorcannotbeopenedeveniftheemergencybuttonisactivated.
Solution:
Pleasecheckiftheairfilterattherearsideofsterilizerisblocked
Takeoffsidecover,unplugthesiliconetubeconnectedtotheairfiltersolenoidvalveanddischarge
thenegativepressureinsidethechamber.
Whenyouheartheexhaustsoundgraduallyweakenuntilitdisappears,youcanopenthedoor.