- 28 -
8 SmallSteamSterilizerMaintenance
8.1 Cleaningofloadeditemspriortosterilization
Beforesterilization,theloadofhandpiecemustbethoroughlycleared.
Dustandcontaminantsseparatedfromtheloadwillcauseblockageoffilters,valvesand
piping,therefore,thesterilizationloadwithoutcleaningisnotallowedtoenterthedevice
forsterilization.
Followallregulationsforcleaning,disinfection,andsterilizationprovidedbythedevice
manufacturer.
8.2 Smallsteamsterilizermaintenance
8.2.1 Cleaningandmaintenance
Regularcleaningandmaintenancecaneffectivelyextendthelifeofthesterilizerandensureagoodworking
condition.
Maintenanceitem Maintenancecycle Specialtool Maintenanceoperation
Replacetheseal Sterilize1000times Openthedoorandcheckwhetherthesealis
deformedorwrinkle,etc.Ifitisdeformedor
wrinkled,itisrecommendedtoreplaceit.
Sealmaintenance 2weeks Removetheseal.Pleaserinseitwithwater
and wipe it dry with a soft cloth.After that
replacethelid.
Note:Payattentiontotheinstallation
directionoftheseal(thenarrowsideis
closetothechambercoverandthewide
sideisclosetothechamber).
Doorlockmaintenance Sterilize1000times lubricatingoil Asmallamountoflubricatingoilissprayed
ontheleftsideofthelidthroughthehole(A
indicated),andtheoutersurfaceiswipedby
apapertowel.
Cleaning distilled water
tank
Sterilize300times
Washitwithcleanwater.
Clean the filter at both
ends of cooling water
inletpipe.
Sterilize100times
Unscrew the nut at both ends of the inlet
pipeandrinseitwithcleanwater.
A