- 17 -
Fig4.3Manualstop
4.6 Sterilizationprocedurecompleted
Whenthesterilizationprocessendsnormally,theprogramendinterfaceandprogramrunningsuccess
informationaredisplayedonthescreen.Iftheautomaticprintingfunctionissetandtheprinterisconnected,the
productwillautomaticallyprintthesterilizationrecord.Pleasepress toopenthedoor.
Carefulburns!
Do not forcethedoortoopen.Failuretodosomayresultin
damage to the product and/or hot steam coming out of the
chamber.
Afterthesterilizationisfinished,waitforawhileuntilthesterilizer
isfullycooledbeforeopeningthedoor.
Donottouchanymetalsurface.
Donottouchthesterilizedloadeditems,chamberanddoorcovers
withbarehandstoavoidbeingburned.
Besuretousethetrayhandletoremovetheloadeditems.
Balancewhenremoving thetraytopreventthe contentsof the
trayfromslipping.
Damagedpackagingdevicesoritemsmaynotbeeffectivelysterilized,
andexposuretothese items mayendangerthe healthofthe operatoror
patient.
Ifdamagedpackagingiffound,pleaserepackagetheitem.
Waiting for 5 minutes after sterilization before removing the
loadeditems,whichisgoodfordryingthesterileitems.
The 11th edition (2017) of the correct handling method for
medical device cleaning and disinfection issued by the
internationalauthorityAKI,Germany.Itisrecommendedthatthe
wet bag be judged as having water drops remaining on the
packageorinthepackageafter15minutesofsterilization.
Physical parameters should be monitored for each sterilization
cycle,recordandcheckprocessvariables.Whenthesterilization
temperaturesetinthiscycleis121°C,theholdingtimeshouldbe
nolessthan15minutes.Whenthesterilizationtemperaturesetin