19
U93
A2081-2.2 en/de/fr HBM
Parameters ExplanationRights
level
UnitValue
1)
Impedance of each
bridge element
345.0 ohm ID Input resistance according to the
HBM data sheet
Response time 1.0000000
u
sec ID Not relevant to HBM transducers.
Excitation level
(nominal)
5.0 V ID Nominal (rated) excitation voltage
according to the HBM data sheet
Excitation level
(minimum)
0.5 V ID Lower limit for the operating range of
the excitation voltage according to
the HBM data sheet.
Excitation level
(maximum)
12.0 V ID Upper limit for the operating range of
the excitation voltage according to
the HBM data sheet.
Calibration date 1‐Jun‐2006 CAL Date the manufacturing certificate
was created at HBM (or date of last
calibration or entry of TEDS data).
Format: day‐month‐year.
Calibration initials HBM CAL Initials of the calibrator or calibration
laboratory concerned (max. three
characters).
Calibration period
(days)
730 days CAL Time before recalibration, calculated
from the date specified under
Calibration Date.
Measurement location
ID
0 USR Identification number for the
measuring point (assigned by the
user). Possible values: 0 to 2047
(numbers only). The HBM Channel
Comment template is also available
as a supplement.
1)
Typical values for an HBM U93/2kN type force transducer, manufactured on 1.6.2006
The user can also write other templates, such as the Signal Conditioning
template.