66
360093-1/H
Platelet Incubator Operation Manual
10 Return the probe to the probe bracket (if previously removed) and reinstall the probe bracket.
a 
(5 cm) from the end of the bracket.
b 
exposed.
c Install the bracket in the chamber.
d If you moved an agitator to access the probe bracket, move the agitator back to its original
position.
7.8.3 Testing temperature alarms automatically on i.Series models

setpoint is reached. During the test, a LOW ALARM TEST IN PROGRESS message appears on the HOME


rises above the setpoint, the audible and visual alarms clear. An event is added to the event log indicating


When you perform an automatic high alarm test, the Peltier device heats the probe until the high alarm
setpoint is reached. During the test, a HIGH ALARM TEST IN PROGRESS message appears on the
HOME screen. When the temperature reaches the setpoint, an alarm sounds, the temperature reading




When you cancel an automatic test, the message indicating that the test is in progress clears immediately. If

To automatically test the low temperature alarm
1 
2 From the MAIN screen, use the buttons to navigate to and select the System Alarm Test & Status
option. The SYSTEM ALARM TEST & STATUS screen appears.
3 
4 Press the ENTER button. You are returned to the HOME screen. Under the reading for the upper
temperature probe, a LOW ALARM TEST IN PROGRESS message appears. When the test is
complete, the LOW ALARM TEST IN PROGRESS message clears.
5 
Compare this value to the alarm setpoint. If the values do not match, see Section 8, “Understanding
and addressing operational issues”
To automatically test the high temperature alarm
1 Identify the current setting for the high alarm setpoint.
2 From the MAIN screen, use the buttons to navigate to and select the System Alarm Test & Status
option. The SYSTEM ALARM TEST & STATUS screen appears.
3 Press the UP or DOWN button until the Start High Alarm Auto Test option is highlighted.
4 Press the ENTER button. You are returned to the HOME screen. Under the reading for the upper
temperature probe, a HIGH ALARM TEST IN PROGRESS message appears. When the test is
complete, the HIGH ALARM TEST IN PROGRESS message clears.