4.6 Battery Charging
The Digilance utilizes a Lithium Ion Battery Pack. The Battery Pack monitors the conditions and cell output, in
order to protect the Battery Cells from Damage, such as Short Circuit or Excessive Temperatures.
• The Charge Level of the Digilance Battery Pack is show in the Upper-Left Corner of the Display.
• Information on the Status of the Battery Pack can be found in the Main Menu
:
• Exchange the battery when the Battery State of Health is
below 50%.
• Charge the Battery Pack with included Power Supply (12V DC, 2A) by connecting the Power Supply
Plug to the Charging Jack:
• During the Charging Process, the Charge Status can be viewed by pressing
Charging Process:
Displayed when the Power Supply is initially connected to the device
Charging Errors:
Power Supply is connected to the device, but there is no charging current
(Blinking)
Battery Temperature is below charging limit (0°C)
If the Battery is Too Cold, the charge current is used to heat the pack,
charging will initiate when the temperature is above the minimum limit.
Battery Temperature is above charging limit (45°C)
Battery Temperature is above maximum limit (60°C)
There is a problem with the Battery Pack. Check state of health of