26 November 2018 Digilance V page 17 of 21
File name: Digilance V_1.01
Version: 1.01
ID
Configuration Section
Options
Default (EU)
Default (US)
4
Power Management
4.1
(L) (R)
Awake Time
Always 180s
15s
15s
4.2
(L) (R)
Motion Sensitivity
LOW/MID/HIGH
Mid
Mid
4.3
Transport Mode
Yes/No
No
No
5
(W)
Communication
(Wireless Versions Only)
5.1
(R)
Wireless ID
1 39
35
35
5.2
(R)(W)
Wireless Mode
Wireless Off
MC Telegram
Data Telegram
MC Telegram & Data
Telegram
MC Telegram
MC Telegram
5.3
(R)(W)
Data Telegram
Telegram 1
Telegram 2
Telegram 3
Programmable Telegram
Telegram 1
Telegram 3
5.4
(W)
Melt Control ID
Alphanumeric, 6 digits
Production
Serial Number
Production
Serial Number
6
Service
6.1
Device Status
Hardware Status Screen
6.2
Load Standard Parameters
Yes/No
No
No
6.3
Reset All Parameters
Yes/No
No
No
6.4
Temperature Offset
-5.0°C to +5.0°C
0°C
0°C
7
(A)
Administrator
7.1
(A)
Region Preference
EU/US
EU
US
7.2
(A)
Calibrate System
Yes/No
No
No
7.3
(A)
Clear Result History
Yes/No
No
No
7.4
(W)(A)
Laird System ID
No (Automatic) or 1 39
No
No
(R)
Reset to Default by Reset All Parameters
(L)
Reset to Default by Load Standard Parameters
(W)
Wireless Instruments Only
(A)
Administrator Access Only
Default TC Type Table
Part Number
Default TC Type
3102001X
TYPE S
3102002X
TYPE R
3102003X
TYPE B
3102004X
TYPE K
6 Advanced Operation (Wireless)
Wireless Communication Options are only available for device Model DL5W
Information on Wireless Communication Configuration can be found in the Wireless Communication