MAINTENANCE
Servicing
The heating unil is virtually maintenance-free
Lubricate the bearing s of the door and door-handle mechanisms with a little synthetic oil from
time to time.
Mineral oil products must not be used where they may come into contact with plastic
components, as they could cause damage to such parts.
Tests
The following items should be tested at least once a year to ensure that the unit remains
in good working order:
Mechanical (unctions
Operation in accordance with the specified technical data
Electrical system (FRG: UVV VBG 4, OIN VOE 0701 Part 1, E DIN VOE 0702 /11.93)
Safety devices
Repairs
Only approved genuine spare parts and accessories may be used.
The use of other parts may result in unforeseen problems and should be avoided under
all circumstances.
Approved spare parts /accessories
Spare part /accessory
Size / model
Instruction manual
Circuit diagram 1/PE AC, 230 V
Circuit diagram 1/PE AC, 120 V
Wire tray with supporting straps
Seasoned silicone rubber door seal
Stacking frame
Space for back panel
Foot
Clasp for the door latching mechanism
Heraeus order number
UT6P
UT12P
UT20P
SO 043 079
SO 041 987
SO 040 640
SO 042 916
SO 042 499
50 042 906
-
50 042 917
SO 042 484
50 042 907
-
50 042 918
SO 042 351
50 042 908
50 028 137
50 042 541
SO 042 284
Other spare parts and accessories to order
30/34
50 042 750
Issue: May, 1994