TECHNICAL DATA
DATA 1 Model
UT6P UT12P
UT20P
Unit
$Wffl^^&$mSS^S.^ * - ' : ••.;
Dimensions
(W x H x D)
Minimum
clearances to
adjacent
surfaces /
objects
Volumes
Weights
Housing
Chamber
Useful space '
Left
Right
Top
Rear
Chamber
(empty)
Useful space ''
Steam space
Unit
552 x 700 X 685
408 X 459 x 334
326 X 367 x 275
696 x 850 x 685
552x610x334
441 x 488x267
754x910x865
610x672x514
488x537x411
50
SO
300 (SO when connected to an extraction system)
120
64.4
33
approx. 73
approx. 40
80
112
57.6
approx. 131
approx. 55
80
210
107.8
approx. 234
approx. 75
Trays
Max. tray load
Number
Dimensions
Lumped
Surface
Total
Standard / Max.
(WxD):
15
20
50
2/9
362 x 333
15
20
50
2/14
506 x 323
10
15
70
2/16
564 x 503
mm
mm
mm
mm
mm
mm
mm
liter
liter
liter
kg
kg
kg
kg
piece
mm
Operating temperatures
Room temperature + 20 to 250 |'C
Temperature deviations, circulating mode, air flap closed
Spatial"
Temporal '>
at 70 *C
at 150 "C
at 250 *C
at 70 *C
at150*C
at 250 *C
+ 0.6
±18
+ 3.4
+ 0.2
+ 0.4
+ 0.7
+ 0.7
+ 1.9
+ 3.5
+ 0.2
±0.3
+ 0.8
+ 1.5
+ 2.8
+ 6.0
+ O.1
+ 07
+ 0.8
K
K
K
K
K
K
Temperature deviations, fresh-air mode, air flap open
Spatial"
Temporal '
at 70 *C
at150*C
at 250 *C
at 70 *C
at 150 'C
at 250 'C
+ 1.3
+ 3.5
+ 5.3
+ 0.2
+ 0.4
+ 1.0
0.7
+ 2.1
+ 3.5
+ 0.2
+ 0.4
+ 0.8
±1-7
3.1
+ 56
+ 0.2
+ 0.5
+ 1.5
|K_
K
K
K
K
K
Warm-up times, circulating mode (unit empty, air flap closed, to 98 % of the operating temperature)
Operating
temperature
70 *C
150 "C
250 *C
8
19
42
8
12
50
9
20
60
min
min
min
1) in accordance with DIN 12 880 Part 2 /11.78
Issue: May, 1994
50 042 750
31/34

Related product manuals