1 Use according to specification
This device is a medical product (laboratory centrifuge) within the context of the IVD Directive 98/79/EC.
The centrifuge was designed only for the separation of materials or mixtures with a density of no more than 1.2
kg/dm³. In particular, these include samples for preparation of in-vitro diagnostic purposes in human medicine.
The centrifuge is only meant for this purpose.
Another use or one which goes beyond this, is considered to be non-intended. The company Andreas Hettich GmbH
& Co. KG is not liable for damage resulting from this.
Observing all information in the operating instructions and complying with the measures described therein is also a
part of the intended use.
2 Remaining risks
The device is built according to the state-of-the-art and the recognized safety regulations. If used and handled
improperly, there could be life-threatening danger to the user or third parties, or the device could be impaired or there
could be other property damage. The device is only to be used for its intended purpose and only when it is in safe
working condition.
Malfunctions which could affect safety must be corrected immediately.
3 Technical specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model MIKRO 120
Type 1204 1204-01
Mains voltage (± 10%) 208 - 240 V 1 100 - 127 V 1
Mains frequency 50 - 60 Hz 50 - 60 Hz
Connected load 230 VA 230 VA
Current consumption 1.2 A 2.5 A
Max. capacity 24 x 1.5 / 2.0 ml
Allowed density
1.2 kg/dm
3
Speed (RPM) 14000
Force (RCF) 17530
Kinetic energy 4100 Nm
Obligatory inspection (BGR 500) no
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 40°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
ΙΙ
Pollution degree
2
Device protection class
Ι
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference,
Interference immunity
EN / IEC 61326-1,
Class B
FCC Class B
Noise level (dependent on rotor)
61 dB(A)
Dimensions
Width
231 mm
Depth
292 mm
Height
199 mm
Weight approx. 7 kg

Table of Contents

Other manuals for Hettich MIKRO 120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich MIKRO 120 and is the answer not in the manual?

Hettich MIKRO 120 Specifications

General IconGeneral
BrandHettich
ModelMIKRO 120
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals