CLX (4/18) Page 1
REV 6.0
Specifications
±5% of reading or ±0.03 mg/L (PPM) whichever is greater for range of
0-6.0 mg/L(PPM)
±10% of reading from 6.01-10.00 mg/L (PPM)
Adjustable; 110 seconds to 10 minutes (600 seconds)
Note: the system defaults to 2.5 minutes
Multi-Line Liquid Crystal Backlit Display
Two Programmable, 120-240VAC 2A Form C Relay
Powered 4-20 mA, 600 drive, isolated
Bi-directional RS-485 with Modbus
Integral pressure regulator 0.34 bar(5.0 PSI) to 10.3 bar (150 PSI.)
0°C – 40°C (32°F – 104°F)
PVC, Borosilicate Glass, Reslyn (FFKM), Viton
®
(FKM),
Polypropylene, Stainless Steel, Acetal, Noryl
®
, Silicone
0°C – 55°C (32°F – 131°F)
Operating temperature for 30 day reagent life is 0-40°C (32°F – 104°F)
Reagent life will be <30 days if kept at temperatures above 40°C
100 – 240 VAC, 47 – 63 Hz, 150VA
100-240 VAC ±10% serial numbers 201705625 or greater
Double Insulated, Pollution Degree 2, Overvoltage Category II
Not recommended for outdoor use.
Up to 95 % RH (non-condensing)
Regulatory Compliance
And Certifications
CE Approved, Listed to UL 61010-1: 2012 Ed 3
Certified to CAN/CSA-C22.2 #61010-1-12: 2012
EMC to EN61326-1: 2013
3.9 kg (8.6 lbs.) Reagents Shipped Separately
406 mm X 406 mm X 241 mm (16”X 16” X 9 ½”)
2 Years from date of manufacture