CLX (4/18) Page 29
REV 6.0
INLET
CHECK
VALVES
CAP ASSEMBLY "A"
PUMP
HAMMER
AND SPRING
THUMB
SCREW
REAGENT
PUMP
ASSY
INLET
OUTLET
FLOW
CHECK VALVE DETAIL
OPTICS SYSTEM
KNURLED TOP
CUVETTE
RETAINING
O-RING
CAP ASSEMBLY "A"
Figure 6: Reagent Tubing Installation
10.2 Preventative Maintenance Schedule
To ensure the instrument will operate reliably, some sites may wish to implement
preventative maintenance. The proposed schedule is shown below:
Procedure
Maintenance Period
Detail
Change Reagents
Once per month
See replacement parts list
Check Cuvette
Once per month
Clean or replace if needed
Check T-Strainer Screen
Once per month
Clean or replace if needed Cat# 28625S
Replace Pump Tubes
Once every six months
Included with Tubing/ Cuvette Kit
Replace Cap Assemblies
Once per year
Tubing/Cuvette Kit Cat # 09950
Replace Check Valves
Once per year
Check Valve Kit Cat # 25017S
Clean T-Strainer Screen
Once per year
Clean or replace if needed Cat #28625S
Follow the enclosed procedures with any of the kits or parts mentioned above.