194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 13
Introduction,
specifications
Characteristics of the weigh system (scale)
The weigh system of the Hill-Rom® 900 Accella™ bed uses four weighing cells installed in
the bearing structure of the bed. The bed is made up of a mobile structure (the weigh
system) and a fixed structure (the frame).
Features Value
Class of the weigh system (2014/31/EEC) NAWI
a
a. Non-Automatic Weigh Instrument
Class III
Maximum range of the weigh system
250 kg
Minimum range of the weigh system
10 kg
Range
0.5 kg
Conditions required to connect the nurse call system
For more information about the connections required to use the nurse call function,
please refer to the
SideCom® Communication System Design and Application
Manual
(DS059).