194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 93
Decontamination,
Maintenance
Servicing the bed
Safety recommendations
Only facility-authorized personnel should perform maintenance of the Hill-
Rom® 900 Accella™ bed.
Before maintenance or servicing works:
ensure that the bed has been immobilized (if no movements are required),
lock out all electrical functions,
disconnect the bed from the mains if no electrical operations are planned,
secure the sleep surface and take whatever steps are necessary to prevent
any movement,
not to intervene on the device if it is busy.
Any device connected to the (CAN) connector exclusively reserved for
maintenance must meet the IEC 60950-1 requirements.
Never open or pierce an electric actuator.
Contact our after sales service for any specific maintenance problem (e.g.,
blockages, etc.).
Preventive maintenance
A service manual and a catalog of spare parts are supplied on delivery, but can
also be obtained on demand from Hill-Rom After-Sales. Hill-Rom guarantees
that the original functional parts or parts performing equivalent functions will
remain available for 7 years after the corresponding range goes out of
production.
The product design life is validated on 10 years of normal use.
The frequency of inspections must be adapted to the general condition of the
product and it use, for example, if the bed is used by heavy patients. It is the
responsibility of the facility to implement a preventive maintenance program for
the bed's functions under its conditions of use.
The bed and accessories should be inspected at least once a year to keep it in
good condition and working properly.
The following points should be given particular attention:
movement mechanisms and cables (actuators in particular),
locking mechanisms (head section, foot section, thigh section and
AutoContour™
),
the accessory mechanisms,
bed movement and ancillary part bearings,
The condition of the electrical cables (e.g., control pendant, power supply
unit, combined mattress connecting cable) in particular that they are not
crushed or cut and thus could make contact with a metal part,