Page 44 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
Preparing the bed for a reset/tare operation* or initializing the Bed
Exit alerts system*
1. Ensure the bed is on a flat surface and that the frame is horizontal (+/-2°).
2. Check that the transport chock protecting the weigh system have been
r
emoved from their housing.
Hill-Rom recommend to keep the transport chocks. in order to protect the weigh
system, they can be re-used in case of bed transfer to another place or building.
3. Install the mattress, cushions, sheets and blankets, and all other accessories that
must remain on the bed.
The weight of these additional articles must not exceed 65 kg or 45 kg,
depending on the destination of the product and the maximum patient weight
(see “Technical specifications” page 11). No more than 39 kg must be added at a
time.