Air circuit leakage, which may be a
result of leakage of valve or air circuit.
Wrong air pressure, which may be a
result of normal open failure of valve.
Weak signal, which may be a result of
too weak pulse of the measured object or
loosen cuff.
Confirm that the
measurement part is
in good contact with
the cuff and whether
the arm
circumference is
within the measuring
range
Out-of-range, which may be a result of
blood pressure of the measured object
exceeding measurement range
Excessive motion, which may be a result
of motion artifacts or more interferences
contained in signals during measurement
Note to avoid
speaking and
disturbance during
the measurement
process and measure
again
Overpressure in measurement, namely
cuff pressure in adult mode exceeding
290mmHg
Timeout in measurement, namely
measurement time in adult mode
exceeding 120 seconds
1. Touch the
start/stop button
2. Detect whether the
button is insensitive
or stuck
Overpressure protection, which is caused
by cuff pressure exceeding the set
maximum value (290)
Sleeve failure, motor failure
The arm posture is not correct and the
elbow switch is not pressed
Adjust the arm
posture, make sure
the elbow switch is
pressed
Communication failure, handshake
communication failure
Failed to start, no response after sending
measurement, unable to start
measurement