Shenzhen Hingmed Medical instrument Co.,Ltd. (Enterprise Edition) 2020-11-18
Shenzhen Hingmed Medical instrument Co.,Ltd.
Width: 310mm Length: 476mm, Height: 300mm
Protection against
electric shock
5 years after installation
Temperature: +5℃~+40℃, relative humidity: 10%-95%
Temperature -20℃~+55℃, relative humidity not more than 95%
Recommendations
of calibration
*********************************************************************
**The blood pressure values obtained by this instrument are equivalent to that fr
om the auscultatory method, with their error consistent with the requirement spe
cified in YY0667-2008. **
*********************************************************************
2. Precautions for safety use
2.1 Contraindications
No
2.2 Warning
*Be sure to use AC for the power supply voltage.
*Please connect to 3P socket, be sure to use by grounding, otherwise it will be easy to
get an electric shock.
*Please do not measure the arm that is getting intravenous drip or blood transfusion,
otherwise it may cause an accident.
2.3 Precautions
Please use this instrument in the following working environment and storage site, if it
is stored or used out of the specified temperature and humidity range (storage
environment: temperature -20℃ ~ +55℃, relative humidity not more than 95%;
working environment: temperature +5℃~+40℃, relative humidity 10%-95%), the
system may not reach the claimed performance specifications.
*Use the instrument in places without water.
*Don't put it in places with high temperature, moisture, direct sunlight and dust, and
don't put in places with salt and sulfur.
*Put it in stable places without tilt, vibration, impact (including during transportation)
*Don't put it in places where chemicals are stored, and don't put it in places where gas
is easy to generate.
*Patients with anticoagulant disease or blood coagulation disorders may have blood
congestion in the place where the cuff is worn even if the cuff is worn in the correct
position during blood pressure measurement.
*If the cuff fails to inflate within 2.5 minutes, instruct the patient to remove the cuff
manually. Prolonged over-inflation may cause blood blockage and make the patient
feel uncomfortable.