EasyManua.ls Logo

HINGMED ABP-03 - User Manual

HINGMED ABP-03
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
CE0120
Hingmed
Ambulatory Blood Pressure Monitor
User Manual
(Apply toABP-03)
Shenzhen Hingmed Medical Instrument Co., Ltd.
VERA/0
Date: May 15, 2014

Questions and Answers

  • M
    Michael MurrayAug 20, 2025
    What to do if HINGMED ABP-03 shows overpressure?
    • M
      matthew46Aug 20, 2025
      If the HINGMED Blood Pressure Monitor displays an overpressure error, where the cuff pressure exceeds 290mmHg in adult mode, 247mmHg in child mode, or 145mmHg in neonatal mode, press the “start/stop” button to measure again.
  • R
    Rachel JimenezAug 22, 2025
    How to fix system error on HINGMED Blood Pressure Monitor?
    • V
      veronica78Aug 22, 2025
      If your HINGMED Blood Pressure Monitor shows a system error, try restarting it. If the error occurs frequently, professional maintenance may be necessary.
  • S
    Sandra BrownAug 25, 2025
    What to do if HINGMED ABP-03 shows weak signal?
    • C
      cburnsAug 25, 2025
      If the HINGMED Blood Pressure Monitor displays a 'weak signal' message, it might be due to a loose cuff. Check the cuff and tighten it if necessary.
  • A
    Alicia SantiagoAug 26, 2025
    How to fix loose cuff on HINGMED ABP-03?
    • F
      fmaciasAug 26, 2025
      If the HINGMED Blood Pressure Monitor displays a message indicating the cuff is too loose, this may be due to loose winding or a disconnected cuff. Retighten the cuff or maintain proper arm position during inflation.
  • A
    Alexandra HuffAug 28, 2025
    What to do if HINGMED ABP-03 is beyond the limit?
    • A
      Allen PerkinsAug 28, 2025
      If the HINGMED Blood Pressure Monitor shows 'beyond the limit', this may be because the subject's pressure exceeds the device's limitation. Press the “start/stop” button to measure again. If the issue persists, consider using a monitor with a larger range.
  • K
    Kristina WilsonAug 30, 2025
    What does signal saturation mean on HINGMED Blood Pressure Monitor?
    • K
      Kristen PerezAug 30, 2025
      If you receive a 'signal saturation' message on the HINGMED Blood Pressure Monitor, where the signal amplitude is too large due to activity or other reasons, keep quiet and press the “start/stop” button to measure again.
  • B
    brittany75Aug 31, 2025
    Why does HINGMED ABP-03 Blood Pressure Monitor show excessive exercise?
    • A
      Amanda YoungAug 31, 2025
      If the HINGMED Blood Pressure Monitor displays 'Excessive exercise', this may be caused by too much motion or interference. Keep quiet during the measurement and avoid moving the arm with the cuff.
  • T
    Thomas DownsSep 2, 2025
    What does overtime mean on HINGMED ABP-03 Blood Pressure Monitor?
    • J
      Joshua SmithSep 2, 2025
      If the HINGMED Blood Pressure Monitor shows an overtime error, where the measuring time exceeds 120s in adult/child mode or 90s in neonatal mode, keep quiet and press the “start/stop” button to measure again.
  • L
    Lisa PageSep 4, 2025
    How to fix air leakage in HINGMED Blood Pressure Monitor?
    • L
      Lee GonzalesSep 4, 2025
      If the HINGMED Blood Pressure Monitor experiences air leakage, possibly due to valve or gas path leakage, tighten the metal connector and check the cuff. If the leakage persists, contact your agent or dealer.
  • T
    Tanya HallSep 6, 2025
    How to fix air pressure error on HINGMED Blood Pressure Monitor?
    • J
      Jeremy TateSep 6, 2025
      If the HINGMED Blood Pressure Monitor experiences an air pressure error, possibly due to the valve being unable to open, check whether the valve can work normally.

Summary

About this Manual

Manufacturer Information

Details of the manufacturing company.

After-Sales Service Contact

Contact information for post-purchase support.

European Representative Contact

Contact details for the EU representative.

Manual Version Details

Details on the manual's version and issue date.

Chapter 1 Preface

Ambulatory Blood Pressure Monitoring Introduction

Explains the concept and benefits of ABPM.

Hingmed ABPM Device Overview

Highlights the monitor's features and body position information.

Hingmed ABPM Scope of Application

Lists the medical conditions for which the device is used.

Chapter 2 Safety Requirements

Device Intended Use

Specifies the device's primary purpose and user limitations.

Device Contraindications

Lists conditions where the device should not be used.

Safety Warnings and Precautions

Important safety precautions and potential hazards.

Important Notices and Guidelines

Specific notes on battery, training, and measurement.

Chapter 3 Product Introduction

Product Composition and Accessories

Lists the main unit and included accessories.

Device Components and Functions

Identifies and explains the device's parts and displays.

Device Symbols Explanation

Explains the meaning of various symbols used on the device.

Chapter 4 Product Installation and Use

Battery Installation Guide

Step-by-step guide for inserting batteries.

Software Installation Procedure

Instructions for setting up the PC software.

Performing a Blood Pressure Study

Steps for performing a monitoring study with the device.

Cuff Selection and Connection

Guidance on choosing and attaching the blood pressure cuff.

Pre-Study Preparation and Training

Pre-study preparation and patient instructions.

Data Derivation and Analysis

Process for transferring and analyzing collected data.

Chapter 5 Troubleshooting and Maintenance

Device Troubleshooting Guide

Lists common errors and their resolutions.

Device Maintenance Procedures

Procedures for device upkeep and inspection.

Chapter 6 Product Specifications

Chapter 7 EMC Information

Chapter 8 Warranty Card

Limited Liability of Guarantee

Outlines the product warranty terms and conditions.

HINGMED ABP-03 Specifications

General IconGeneral
ModelABP-03
Measurement MethodOscillometric
Accuracy (Pressure)±3 mmHg
DisplayLCD
TypeAutomatic Blood Pressure Monitor
Power Source4 x AA batteries or AC adapter
Operating Humidity15% to 85% RH
Measurement Range (Pulse)40-200 beats/min