20
EC36 Measurement result is not available
Press “start/stop” button to measure
again.
EC37 Overtime (beyond 180s) remeasure
EC48 Memory is full, can no longer measure Upload and empty the data
* If the errors can be revised, please call agent or after-sale department of Hingmed for
help.
5.2 Maintenance
5.2.1 Maintenance inspection and safety management
Visually inspect the broken or damages on tube connecters, pressure air tube and the
enclosure of monitor. If any damage is found, do not use the monitor. Please contact our
after-sale service department.
5.3 Device Maintenances
After use, it is important to perform preventative maintenance to ensure the safe and
efficient operation of the monitor for long-term.
● Do not disinfect. DO NOT immerse the monitor in any fluid, or attempt to clean with any
liquid detergents, cleaning agents, or solvents. You may use a soft, damp cloth to
remove dirt and dust from the monitor. If the unit does become immersed in water, do
not use and contact our service department.
● Do not clean the enclosure and cuff with strong alkali or acid or strong disinfectant.
● May use a mild detergent to clean the cuff, belt and bladder; alternatively, may also
wash them in a washing machine. Remove the bladder from the cuff before machine
washing. Wash them by using warm water and a mild detergent, then hang it for drying.
Any problem, please contact local distributors or Hingmed (see Chapter 8 for
information).
● It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of
the cuff pressure indication and Air leakage (testing at least at 50 mmHg and 200
mmHg).
Chapter 6 Product specifications
Name Ambulatory Blood Pressure Monitor
Model ABP-03
measuring method Oscillometric method