Product composition: Main unit, Noninvasive blood pressure cuff, USB data wire and
Hingmed PC software.
The device is inner power source with 2 AA alkaline batteries; Electric shock protection
degree: CF type applied part; Operation mode: continuous mode.
The service life of this product is five years.
2.2 Contraindication
Do not use the monitor near X-ray tomography device.
Do not use the monitor in the places where inflammable anesthetic exist, it may lead to
explode.
Do not wrap the cuff on the limb being used for IV injection as cuff inflation may block
infusion and hurt patient.
Do not immerse the monitor in any liquid or any detergent, which would cause electrical
hazard.
2.3 Warnings
Cuff shall not be worn on the wound, because inflation pressure may cause further
damage;
When patients carry the monitor, pull out the data wire from PC or other devices;
Never apply the device to patient under 3 years old and patient who can’t express their
thoughts.
The instrument can’t be used in conjunction with defibrillation equipment.
It may cause measurement error by using parts not included in the supply listing.
Shoulder straps or cuff tube winding patient's neck may cause suffocation, doctor
should remind patient to wear the cuff on upper arm, and shoulder straps or cuff tube
must not wind patient's neck. Tube should be placed under the coat (even at night).
Make sure the pressure is suitable for the patient. If any abnormity occurs in the
monitoring process, please stop measuring immediately, and remove the cuff from the
patient or shortly press start/stop button to stop inflating. If the cuff fails to deflate,
indicate patient to remove it properly and safely.
If this device is dampened by accident, put it in a well-ventilated place for drying prior to
use.
Only professional physicians can explain the measured blood pressure values.
Only medical personnel and trained professionals can use the device.
Dont servicing/maintenance while the medical equipment is in use
The patient is an intended operator. The patient can only measure, transmit data and
charge battery under normal circumstances and maintain the device and its
accessories according to the User Manual.
Not intended to be sterilized.
Not for use in an oxygen rich environment.
No modification of this equipment is allowed.
It is not intended for use with neonate.
It is not intended for use with pregnant or pre-eclamptic patients.