Shenzhen Hingmed Medical instrument Co.,Ltd. (Enterprise Edition) 2020-11-18
Shenzhen Hingmed Medical instrument Co.,Ltd.
EC36
No measurements. The instrument cannot
get measurement results.
remeasurement
EC37
180S timeout
remeasurement
EC64
Out of paper in printer
Follow the manual to
replace the printing
paper
EC65
The printer is not closed, and the printer
cover is not closed.
Check whether the
printer cover is
closed.
9. Maintenance
************************************************************
Note: *The sphygmomanometer and accessories do not need to be sterilized, but
should be kept clean. If there is contamination, it should be cleaned and disinfected in
time. In order to avoid long-term damage to the products, we suggest you sterilize
them only if necessary according to regulations of your hospital.
*After use for infected people or people with suspected infection, the contact parts
with the patients should be disinfected
*When cleaning and disinfecting, do not soak the product and accessories in liquid.
Do not allow liquid to flow into the connection socket or case of the
sphygmomanometer to prevent damage to the sphygmomanometer.
************************************************************
9.1 Cleaning
(1) Before cleaning the sphygmomanometer, turn off the power supply of the host and
disconnect the AC power supply.
(2) Wet the soft and clean lint-free cloth with mild soapy water or non-corrosive
diluted detergent.
(3) Wipe the contact surface between the instrument and the patients.
(4) Dry with a clean and dry soft cloth.
(5) Blood pressure cuff: After soaking in soapy water, rinse and dry
9.2 Disinfection
It is recommended that users use 70%~80% (volume ratio) ethanol disinfectant to
soak a piece of clean dry gauze, and then wipe the surface of the object to be
disinfected with the gauze twice for 3 minutes. Air dry or wipe off the residual
disinfectant with a clean and dry cloth.
*******************************************************
Note: Clean before disinfection. Please keep away from fire during the disinfection
process because ethanol is flammable. People who are allergic to alcohol should use
ethanol disinfectant with caution.
*******************************************************
9.3 Maintenance of print head
(1) Turn off power supply of the instrument.
(2) Press the printer cover plate button to open the cover plate.
(3) Use a cotton swab dipped in alcohol or a soft cotton cloth to wipe the print head
from the head to the tail in the same direction.
(4) Please wipe off rubbish, dust, paper scraps and other foreign objects in the printing
paper storage box.