Shenzhen Hingmed Medical instrument Co.,Ltd. (Enterprise Edition) 2020-11-18
Shenzhen Hingmed Medical instrument Co.,Ltd.
we recommend to use
UPS or battery for
power supply.
Power frequency
magnetic field
(PFMF)
(50Hz/60Hz)G
B/T 17626.8
PFMF should possess
the horizontal
characteristics of
power frequency
magnetic fields in
typical places in typical
commercial or hospital
environments.
Note: UT means AC network voltage before applying test voltage.
Guidelines and Statement of Manufacturer - Electromagnetic Immunity
The instrument is expected to be used in following specific electromagnetic environments, and
the purchaser or the user should ensure that it is used in these electromagnetic environments:
Electromagnetic
Environment - Guidelines
3V (effective value)
150kHz~80MHz
The distance between
portable or mobile RF
communication device and
any used part (including
cable) of DBP-01H should
not less than the
recommended isolation
distance. The distance
should be calculated by the
formula corresponding to
the transmitter frequency.
Recommended isolation
distance
P - the maximum rated
output power of transmitter
provided by the transmitter
manufacturer, in watt (W);
d - the recommended
isolation distance in meters
(m)
The field intensity of
stationary RF transmitter is
determined by the
electromagnetic field
survey a, which should be
lower than the conforming
level at each frequency