10
Safety Information
Character expression
The instrument’s display expresses the alphanumeric characters as
follows.
1 2 3 4 5 6 7 8 9 0
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Some di󰀨erent expressions are used as below:
Indicates that the saved data has been deleted.
Indicates that the comparator buzzer is set to
FAIL.
Indicates that the A/D converter communications
error occurs.
Accuracy labeling
The instrument accuracy is expressed by dening a percentage of
the reading and a limit value for errors in terms of digits.
reading Display value
Indicates the value displayed by the instrument. Limit values
for reading errors are expressed as a percentage of the
reading (“% of reading”).
digits Resolution
Indicates the minimum display unit (in other words, the
smallest digit that can have a value of 1) for a digital
measuring instrument. Limit values for digit errors are
expressed using digits.