EasyManua.ls Logo

Hioki CM4371 - Page 55

Hioki CM4371
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52
Accuracy Table
AC current (PEAK MAX/PEAK MIN)
Range
(Accuracy guarantee
range)
Resolution
Accuracy guarantee
frequency range
Measurement accuracy
600.0 A (±10 A to ±1500 A) 1 A 10 Hz≤f<45 Hz ±1.8% rdg.±7 A
45 Hz≤f≤66 Hz ±1.3% rdg.±7 A
66 Hz<f≤1 kHz ±2.0% rdg.±7 A
2000 A (±10 A to ±2300 A) 1 A 10 Hz≤f<45 Hz ±1.8% rdg.±7 A
45 Hz≤f≤66 Hz ±1.3% rdg.±7 A
66 Hz<f≤1 kHz ±2.0% rdg.±7 A
(±2301 A to ±2840 A) 1 A 10 Hz≤f<45 Hz ±6.5% rdg.±7 A
45 Hz≤f≤66 Hz ±6.0% rdg.±7 A
66 Hz<f≤1 kHz -
6
DC current and auto A DC detection 600.0 A/2000 A (CM4373, CM4374)
Conditions of guaranteed accuracy: After zero adjustment has been performed
Zero-display range: 5 counts or less
Coupling type: DC coupling
Peak detection time width: 1 ms or more (Filter off)