Accuracy Table
AC+DC voltage (PEAK MAX/PEAK MIN)
Range
(Accuracy guarantee
range)
Resolution
Accuracy guarantee
frequency range*
1
Measurement accuracy
6.000 V (0.00 V to ±12.00 V) 0.01 V 10 Hz≤f<45 Hz ±1.5% rdg.±0.07 V
DC, 45 Hz≤f≤66 Hz ±1.0% rdg.±0.07 V
66 Hz<f≤1 kHz ±1.5% rdg.±0.07 V
60.00 V (±3.0 V to ±120.0 V) 0.1 V 10 Hz≤f<45 Hz ±1.5% rdg.±0.7 V
DC, 45 Hz≤f≤66 Hz ±1.0% rdg.±0.7 V
66 Hz<f≤1 kHz ±1.5% rdg.±0.7 V
600.0 V*
2
(±30 V to ±1000 V) 1 V 10 Hz≤f<45 Hz ±1.5% rdg.±7 V
DC, 45 Hz≤f≤66 Hz ±1.0% rdg.±7 V
66 Hz<f≤1 kHz ±1.5% rdg.±7 V
1000 V*
3
(±50 V to ±1000 V) 1 V 10 Hz≤f<45 Hz ±1.5% rdg.±7 V
DC, 45 Hz≤f≤66 Hz ±1.0% rdg.±7 V
66 Hz<f≤1 kHz ±1.5% rdg.±7 V
*1 Frequency range of 10 Hz≤f<20 Hz is designed value.
*2 Values of up to ±1200 V are displayed, but accuracy is not dened for display values in excess of 1000 V
(which are provided as reference values).
*3 Values of up to ±1700 V are displayed, but accuracy is not dened for display values in excess of 1000 V
(which are provided as reference values).