Accuracy Table
14
Resistance measurement (same for all models)
Maximum capacity load: 10 mF
Maximum inductive load: 10 H
Conditions of guaranteed accuracy: After zero adjustment has been performed
Overload protection: Lower of 1700 V DC/1000 V AC or 2×10
7
V • Hz (energized for 1 minute)
Overload current: 30 mA or less at steady state, 1.5 A or less at transient state
Resistance measurement (Measurement value/MAX/MIN/AVG)
Range
(Accuracy guarantee
range)
Resolution
Measurement
current
Measurement accuracy
Open terminal
voltage
600.0
Ω
(0.0
Ω
to 600.0
Ω
) 0.1
Ω
200 µA±20% ±0.7% rdg.±0.5
Ω
2.0 V DC or
less
6.000 k
Ω
(0.000 k
Ω
to 6.000 k
Ω
) 0.001 k
Ω
100 µA±20% ±0.7% rdg.±0.005 k
Ω
2.0 V DC or
less
60.00 k
Ω
(0.00 k
Ω
to 60.00 k
Ω
) 0.01 k
Ω
10 µA±20% ±0.7% rdg.±0.05 k
Ω
2.0 V DC or
less
600.0 k
Ω
(0.0 k
Ω
to 600.0 k
Ω
) 0.1 k
Ω
1 µA±20% ±0.7% rdg.±0.5 k
Ω
2.0 V DC or
less
Auto range movement threshold: 6000 counts or more for upper range, 540 counts or less for lower range.