69
Accuracy Table
19
DC power 600.0 A (CM4371, CM4372)
Displays the product of DC current and DC voltage.
Accuracy guarantee conditions, zero-display range, connection method, and band depend on the DC current
and DC voltage.
Display range switching* Minimum resolution Measurement accuracy
0.0 VA to 1020 kVA 0.1 VA ±2.0% rdg.±20 dgt.
* Switches the display range automatically based on the voltage range.
20
DC power 2000 A (CM4373, CM4374)
Displays the product of DC current and DC voltage.
Accuracy guarantee conditions, zero-display range, connection method, and band depend on the DC current
and DC voltage.
Display range switching* Minimum resolution Measurement accuracy
0.000 kVA to 3400 kVA 1 VA ±2.0% rdg.±20 dgt.
* Switches the display range automatically based on the voltage range.