49
Accuracy Table
索
引
6
5
4
3
2
1
(9) DC voltage
Measurement value/MAX/MIN/AVE
Range (Auto-range threshold)
Accuracy guarantee range
(Resolution)
Measurement
accuracy
Input impedance (DC
input)
600.0 mV
(more than 6000 counts)
0.0 mV to ±600.0 mV
(0.1 mV)
±0.5% rdg.
±0.5 mV
6.7 M
Ω
±5%
6.000 V (more than 6000 counts/
less than 540 counts)
0.000 V to ±6.000 V
(0.001 V)
±0.5% rdg.
±0.003 V
6.7 M
Ω
±5%
60.00 V (more than 6000 counts/
less than 540 counts)
0.00 V to ±60.00 V
(0.01 V)
±0.5% rdg.
±0.03 V
6.1 M
Ω
±5%
600.0 V (more than 6000 counts/
less than 540 counts)
0.0 V to ±600.0 V
(0.1 V)
±0.5% rdg.
±0.3 V
6.0 M
Ω
±5%
1500 V (less than 540 counts) 0 V to ±1000 V *
1
(1 V)
±0.5% rdg.
±3 V
6.0 M
Ω
±5%
±1001 V to ±1700 V *
1
(1 V)
±2.0% rdg.
±5 V
*1: In the 1500 V range, the instrument can withstand input of up to 1000 V continuously or input in excess of
1000 V for no greater than 1 minute.