35
Input specications/Measurement specications
索
引
6
5
4
3
2
1
3.2 Input specications/Measurement specications
(1) Basic Specications
Measurement range See "3.3 Accuracy Table" (p. 42)
Maximum rated voltage
to terminal
1000 V AC (up to 1 kHz)
1700 V DC
Maximum rated voltage
to earth
600 V AC (Measurement category IV)
1000 V AC (Measurement category III)
Anticipated transient overvoltage: 8000 V
Measurement method True RMS measurement
Measurement terminals COM terminal and V terminal
(2) Current measurement specications
Maximum input current See the frequency derating characteristics (p. 37)
Coupling type AC current *
1
AC coupling
Other current measurements than
the above
DC coupling