54
Accuracy Table
(11) Continuity check
Range Accuracy guarantee range (Resolution) Measurement current Measurement accuracy
600.0
Ω
0.0
Ω
to 600.0
Ω
(0.1
Ω
) 200 µA±20% ±0.7% rdg.±0.5
Ω
(12) Resistance
Range (Auto-range threshold)
Accuracy guarantee
range (Resolution)
Measurement
current
Measurement
accuracy
600.0
Ω
(more than 6000 counts)
0.0
Ω
to 600.0
Ω
(0.1
Ω
)
200 µA±20% ±0.7% rdg.
±0.5
Ω
6.000 k
Ω
(more than 6000 counts/less than 540 counts)
0.000 k
Ω
to 6.000 k
Ω
(0.001 k
Ω
)
100 µA±20% ±0.7% rdg.
±0.005 k
Ω
60.00 k
Ω
(more than 6000 counts/less than 540 counts)
0.00 k
Ω
to 60.00 k
Ω
(0.01 k
Ω
)
10 µA±20% ±0.7% rdg.
±0.05 k
Ω
600.0 k
Ω
(less than 540 counts)
0.0 k
Ω
to 600.0 k
Ω
(0.1 k
Ω
)
1 µA±20% ±0.7% rdg.
±0.5 k
Ω
(13) Diode
Range Accuracy guarantee range (Resolution) Short-circuit current Measurement accuracy
1.800 V 0.000 V to 1.800 V *
1
(0.001 V) 200 µA±20% ±0.7% rdg.±0.005 V
*1: Beeping buzzer tone at forward connection (0.15 V to 1.8 V). Continuous buzzer tone and red backlight
lights up if less than 0.15 V.