DT4281,DT4282 DIGITAL MULTIMETER Remote Operation Manual
P6/14
5. Detail of Commands
5.1. Description of Commands
The DT4281,DT4282 only accepts the UPPER CASE command. (Except Unit etc.)
Table.4
Explanation
Query the Meter model.
Syntax QPID
Response DT4281 or DT4282
Query the Meter identification.
Syntax *IDN?
Response <data×3>
<data> maker name, model number, serial, version
Example; “HIOKI,DT4281,121107517,Ver 1.00”
Put the meter to reset.
Syntax :SYST:RST
Response “OK”
Put the meter into the local lockout state.
Syntax :SYST:LLO
Response “OK”
Put the meter into the local state.
Syntax :SYST:GTL
Response “OK”
Query the function of the display.
Syntax :CONF? [:CONF2?]
Response <data×2>
<data> function,range
Example; “ACV, 600m”
5.2 Combination response string function and range
Configure the range.
Syntax :CONF <data×2>
<data> function,range
Example; “:CONF RES, 60k”
5.2 Combination response string function and range
Response “OK” or “CMD ERR” or “EXE ERR”