22
Instrument Overview and Features
Maintain the sampling speed, even when modules are added.
Adding modules, each of which incorporates an A/D converter, does not cause the reduction
of the sampling speed.
You can perform measurement at di󰀨erent data refresh rates (sampling intervals) appropriate
for every module.
See “Measurement module data refresh intervals” in the Instruction Manual.
Setting example
Control signals are recorded at a high sampling
interval
of 1 ms using the U8553 High
Speed Voltage Unit.
Multi channels of temperatures are recorded at a sampling
interval
of 1 s using the U8552
Voltage/Temp Unit.
(The U8552 can use a relatively low sampling speed, allowing the lter to set a lower cuto󰀨
frequency. Using the U8552 can eliminate lower-frequency noise.)
Perform strain measurement.
You can measure multiple channels of strain data using the Strain Unit.
The instrument has a built-in bridge box, allowing it to support a variety of connection
methods.
See “Connecting a strain gage or converter” (p. 58).