EasyManua.ls Logo

Hioki LR8450 - Setting Example; (Measuring Temperature Using Thermocouples

Hioki LR8450
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
83
Setting Example (Measuring Temperature Using Thermocouples)
3.2 Setting Example
(Measuring Temperature Using Thermocouples)
This section describes how to congure the settings to measure temperature with the instrument
and a plug-in module
You can measure temperature with the following products.
Logger: LR8450 or LR8450-01
Plug-in module: U8550 Voltage/Temp Unit
Thermocouple: Type K
Preparing for measurement
1
Insert storage media.
See “2.7 SD Memory Card and USB Drive” (p. 68).
2
Connect the thermocouple (Type K) to the U1-1 input terminal of the module.
Cable insulation colors vary by country and manufacturer. Check before connecting.
See “Connecting voltage cables and thermocouples” (p. 52).
How to operate the instrument
1
Press the SET key to display the [Measure] main tab and the [Record] sub tab.
Set the recording conditions as follows:
Main tab
Sub tab
11
33
22
44
55
No. Setting Condition Description
1 Recording mode Normal Fixed.
2 Recording interval 1 s Measures temperatures every 1 s.
3 Repetitive recording OFF Disables the repetitive recording. The instrument will
perform a series of measurements only once.
4 Recording time Continuous Continues the recording until you press the STOP key.
5 Alarm source OFF Disables the alarm function.
Settings and Operation