87
4
Specications
For detailed information on the instrument’s specications, see “10 Specications” in the LR8450/
LR8450-01 Instruction Manual stored on the accompanying CD.
4.1 Memory HiLogger Basic Specications
Operating
environment
Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Operating
temperature and
humidity range
−10°C to 50°C (14°F to 122°F), 80% RH or less (non-condensing)
(Charging temperature range: 5°C to 35°C [41°F to 95°F])
Storage temperature
and humidity range
−20°C to 60°C (−4°F to 140°F), 80% RH or less (non-condensing)
Power supply AC adapter Z1014 AC Adapter (12 V DC ±10%)
AC Adapter rated supply voltage
100 V to 240 V AC (assuming voltage uctuation of
±10%)
AC Adapter rated power supply frequency
50 Hz/60 Hz
Battery LR8450 accommodates 2 batteries.
Z1007 Battery Pack
(When used with AC Adapter, AC Adapter has priority.)
Li-ion 7.2 V, 2170 mAh
External power supply 10 V to 30 V DC
Power consumption Normal power
consumption
Using the Z1014 AC Adapter or 12 V DC external power supply,
without Battery Pack
With LCD at maximum brightness: 8.5 VA (instrument only)
With LCD backlight o󰀨: 7 VA (instrument only)
Maximum rated power When using the Z1014 AC Adapter
95 VA (including AC Adapter)
When using a 30 V DC external power supply
28 VA (while charging battery with LCD at maximum
brightness)
When using the Z1007 Battery Pack
20 VA (with LCD at maximum brightness)
4 Specications
Specications
HIOKI LR8450A963-02

Table of Contents

Other manuals for Hioki LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki LR8450 and is the answer not in the manual?

Hioki LR8450 Specifications

General IconGeneral
ModelLR8450
CategoryData Loggers
Dimensions330 mm (12.99 in) W × 221 mm (8.70 in) H × 120 mm (4.72 in) D
WeightApprox. 2.8 kg (excluding battery)
Operating Temperature0 to 40 °C (32 to 104 °F), 80% RH or less (no condensation)
Sampling Rate1 ms to 1 hr
MemoryInternal memory
InterfaceUSB, Ethernet
Power SupplyAC adapter or battery
Measurement RangeVaries with module
AccuracyVaries with module
Storage MediaSD card
Memory CapacityInternal memory
Input TypeVoltage, temperature
Temperature MeasurementVaries with module
Communication InterfaceUSB, Ethernet
Measurement ChannelsUp to 120 channels (with optional input modules)