84
Setting Example(Measuring Temperature Using Thermocouples)
2
Press the SET key to display the [Channel] main tab and the [Individual] sub tab.
The instrument represents the input channels as Un-m.
n represents the module number; m represents the channel number.
Set the input conditions of U1-1 (Channel 1 of Module 1) as follows:
Main tab
Sub tab
11
33
22
44
55
66
Channel U1-1
No. Setting Condition Description
1 Check box
Performs measurement for channels whose check box
is selected.
2 Input type Thermocouple Measures temperature using thermocouples.
3 Range 100°C Measures temperature between −100°C and 100°C.
4 Type K Select Type K thermocouple.
5 Burn out OFF Disables the thermocouple wire break detection
function.
6 RJC INT Performs the reference junction compensation in the
measurement module.
Switching over to the [Unit 1] sub tab allows you to congure the settings for 15 channels at once.
Clear the check boxes of channels U1-2 to U1-15, exempting the channels from the measurement
subjects.