APT4000PH
Configuration
Output 1
Output current during Error and Hold.
Select measured variable
Select electrode type
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
End:
Press CONF, then ENTER
Code
Output 1
o1.
Display
Menu group
Select menu item
39
OFF
(ON)
Code
Display
Action
Choices
o1.
22 mA signal for error message
Select with key
Proceed with ENTER
Output signal during Hold
LAST:
During Hold the last measured value
is maintained at the output
FIX:
During Hold a value (to be entered) is
maintained at the output.
Select with key
Proceed with ENTER
LAST
(FIX)
Only with FIX selected:
Enter current which is to flow at the
output during Hold
Select with key,
edit number with key,
proceed with ENTER
021.0 mA
(000.0 to
021.0 mA)
Output signal for Hold:
Output current
[mA]
Output signal Hold
Setting FIX = 21.0 mA
Hold active
21
4
Output signal HOLD
Setting LAST
Hold active